Q:

Which statements are correct comparisons of gated equilibrium and initial-transit radionuclide measurements of left ventricular function?

0

Which statements are correct comparisons of gated equilibrium and initial-transit radionuclide measurements of left ventricular function?


  1. Gated equilibrium techniques provide more accurate measurements of ejection fraction than initial-transit methods
  2. Left ventricular imaging time for a gated equilibrium study is at least 10 times that of an initial-transit study.
  3. Both techniques require the same radiopharmaceuticals.
  4. Both techniques require a bolus injection.

All Answers

need an explanation for this answer? contact us directly to get an explanation for this answer

B. Left ventricular imaging time for a gated equilibrium study is at least 10 times that of an initial-transit study.

DISCUSSION: Both techniques are equally accurate for measuring left ventricular ejection fraction. The left ventricular imaging time for gated equilibrium studies is at least 10 times that of initial-transit radionuclide angiocardiography. Initialtransit techniques use data from fewer than 10 heartbeats, whereas equilibrium studies require more than 100 heartbeats to acquire data with similar information density. The initial-transit study can be performed with any radioactive substance, but the gated equilibrium technique requires a radiopharmaceutical that remains within the blood pool for imaging. The initial-transit radionuclide study requires a bolus injection, but an equilibrium study can be acquired up to several hours after injection and must be acquired while the tracer is at equilibrium. 

need an explanation for this answer? contact us directly to get an explanation for this answer

total answers (1)

Similar questions


need a help?


find thousands of online teachers now